Puno Filter

Puno Filter / Guides / Filter integrity testing

Filter integrity testing: bubble point, diffusion and water intrusion

An integrity test shows that a membrane filter has no defect and the pore size it claims, without destroying it. Hydrophilic cartridges are tested by bubble point or diffusion flow; hydrophobic vent cartridges are usually tested by water intrusion.

The four tests

TestHow it worksUsed onA fail means
Bubble pointThe membrane is wetted, then gas pressure is raised until gas first pushes through the largest poresHydrophilic membranes wetted with water; hydrophobic membranes wetted with alcoholA defect, the wrong pore size, or incomplete wetting
Diffusion (forward flow)Gas is held at a set pressure below the bubble point and the small flow through the wetted membrane is measuredLarger cartridges and multi-cartridge housingsA defect or a damaged seal
Pressure holdThe upstream side is pressurised and the drop in pressure over time is recordedInstalled filters, as a simpler form of the diffusion testA leak in the filter or the housing
Water intrusionWater is pressed against a dry hydrophobic membrane and the tiny flow into it is measuredHydrophobic PTFE vent and gas filtersA defect, or a membrane that has lost its water repellence

Why hydrophobic filters get a different test

A hydrophobic membrane does not wet with water, so a water-wetted bubble point is not possible. Wetting it with alcohol works, but the alcohol then has to be flushed and dried out of a vent filter. The water intrusion test avoids that: the filter stays dry and nothing but water touches it.

When filters are tested

Which of these your plant needs depends on the rules it works under.

What a test does not tell you

An integrity test shows the filter is intact. It does not show the filter suits your liquid. Limits for the test come from the maker and apply to a stated wetting liquid and temperature; a liquid other than water changes them.

How PUNO cartridges are tested

Every absolute-rated PUNO membrane cartridge is integrity tested at the factory by bubble point, diffusion flow or water intrusion, and is individually numbered with batch records. You can ask for the record with your order. See quality and documents, or the PES, PVDF and PTFE cartridge pages.

Published . Last updated .

Request a quote

The more you can tell us about the duty, the closer the first quote will be. A photo or model number of your current cartridge is enough to start.

Your message goes to sales@punofilter.com and we reply by email.